PCL120 
Very High Accuracy: ±0.015% of Reading 
Very High Resolution: 0.001 mA 
Auto Step and Ramp Function
% Error Function— Eliminates Manual Error Calculation 
Extended Adjustment Range—Digital Knob and User-Selectable Decade Output Allow Large and Incremental (0.001 mA Step) Output Changes 
Reads 0 to 28 Vdc 
Built-In 250 Ohm Resistor Facilitates Calibration of Hart? Devices 
Remote Calibration Through Automatic Step and Ramp Functions 
Loop Power Measurement Capability 
Simulates, Powers, and Measures 2-Wire Transmitters 
5 Preset Outputs (4, 8, 12, 16, 20 mA) 
Fuseless Input Protection Up to 250 Vac 
With an accuracy of 0.015% of reading, 0.001 mA resolution, and 1 microamp step resolution, the PCL120 has the highest precision in its class. Other features that make the PCL120 stand out include a "% error" function, which eliminates manual error calculation and allows the display of actual versus ideal error at any calibration point. The PCL120 can simulate, power, and measure 2-wire transmitters. Its automatic step and ramp functions enable remote calibration of 4 to 20 mA devices.
The PCL120 has many innovative, powerful features that make calibration easy. A few clearly marked keys give the user access to the full functionality of the unit. The STEP and RAMP output modes automate either a discrete 4-8-12-16-20-16-12-8-4 mA stair step output or 2 speeds of a linear 4-20-4 mA ramp output. Pressing the AUTO key (1) activates this time-saving function. To further aid remote calibration, the PCL120 features a minimum/maximum value recall function. In the measure mode, holding the AUTO(/MAX) key (1) gives a direct readout of the maximum value, while holding the 25%(/MIN) key (4) gives a direct readout of the minimum value. Holding the % ERROR key (2) while turning the large knob provides instant feedback on the difference between the ideal and actual current values required for a given device response. This ls the user whether the device is within its accuracy range. No further calculations are necessary. A LOOP POWER key (3) provides instant 24 Vdc loop power output. The 25% (4) key provides instant outputs of 4, 8, 12, 16, and 20 mA loop current, with increments occurring each time the key is pushed.
SPECIFICATIONS (25°C Unless Otherwise Noted) 
Input 
Current Range: 0 to 24 mA; -25 to 125% 
Voltage (Reading): 0 to 28 Vdc 
Input Protection: Fuseless up to 250 Vac 
Output 
Source/Simulate Range: 0 to 24 mA; -25 to 125% 
Drive Capability: 
Without HartTM: 1200 O 
With HartTM: 950 O 
Loop Supply Voltage: 24 Vdc 
Range Select: Decade; incremental in 0.001 mA steps 
Resolution: 1 μA 
Accuracy: ±0.015% of reading ±2 μA 
Environmental
Operating Temperature: -10 to 55°C (14 to 131oF) 
Storage Temperature: -20 to 70°C (-4 to 158oF) 
Power Requirement: 9 Vdc 
Battery: One 9 V (included) 
AC Adaptor: Optional 
Mechanical Dimensions (Without Boot): 144.7 H x 80 W x 3.63 mm D (5.7 x 3.15 x 1.43") 
Weight: 340 g (12 oz)
 
产品型号 产品描述
PCL120 Precision loop calibrator
PCL-ACA 120 Vac power adaptor (PCL422A-BATT required)
PCL422A-BATT 9 V NiCad battery 